SNSPD
Cryostat
We provide a low-temperature niobium-based SQUID magnetometer developed by SIMIT.
Application scope: Biomagnetic detection, geophysics, three-axis compensation, etc.
User customization: A variety of SQUID magnetometers can be designed to meet different application needs.
Packaging design: reliable packaging of non-magnetic materials, such as glass fiber reinforced plastic, quartz, etc.
Working conditions: The magnetometer can be directly immersed in liquid helium.
Signal readout: used for self-made readout circuit to achieve amplification and feedback, that is, to achieve SQUID flux locking loop readout
Type | Type A | Type B | Type C | |
Chip size (mm2) | Type A-1:10×10 | Type A-2: 5×5 | 5 × 5 | 5 × 3 |
Magnetic field sensitivity (nT/Ф 0) | 0.58 | 1.8 | 1 | 69 |
Magnetic flux noise (µ ф0/√Hz) | 5 | 5 | 7 | 10 |
Magnetic field noise (fT/√Hz) | 2.9 | 9 | 7 | 690 |
Squid configuration | One step degree | Multi-loop | Bare washer | |
Inductance (pH) | 350 | 350 | 320 | 120 |
Detection coil size (mm 2) | 9.4 × 9.4 | 4.7 × 4.7 | - | - |
Heating resistance (Ω) | 100 | 100 | 50 | 100 |
Critical current (µ a) | 18 | 18 | 10 | 18 |
Junction shunt resistance (Ω) | 12 | 12 | 10 | 7 |
Junction capacitance (pH) | 0.45 | 0.45 | 0.45 | 0.45 |
Β l shielding coefficient | 3 |